GDC-9560B Power System Insulation Oil Gas Chromatograph Analyzer
ʻO ka GDC-9560B Gas Chromatograph Analyzer ka nānā ʻana i ka ʻike kinoea o ka ʻaila insulation ma ke ʻano chromatographic gas.He mea maikaʻi ke hoʻoholo inā he hewa paha ka mea holo e like me ka wela nui, hoʻokuʻu a ʻaʻole paha, e hōʻoia i ka palekana o ka mana mana.Pono nō hoʻi e hoʻohana ʻia no ka mea hana i nā mea hana kinoea/ʻaila e nānā i kā lākou mau lako ma mua o ka haʻalele ʻana i ka hale hana.
ʻO ka ʻenehana loiloi chromatography gas kahi ʻenehana no ka hoʻokaʻawale ʻana a me ka nānā ʻana o nā hui ʻana o nā mea he nui.Hoʻohana nui ia i ka ʻokoʻa o ka paila paila, polarity a me adsorption coefficient o kēlā me kēia ʻāpana i ka hāpana i loko o ke kolamu, no laila ua hoʻokaʻawale ʻia nā mea i loko o ke kolamu, a ʻo nā mea i hoʻokaʻawale ʻia e qualitatively a quantitatively analysisd.
Hoʻohana ka chromatograph kinoea i ke kinoea ma ke ʻano o ke kaʻa kaʻa (keea lawe lima).Ke hoʻouna ʻia ka hāpana i ka injector a vaporized, lawe ʻia ia e ka mea lawe kinoea i loko o ke kolamu i hoʻopaʻa ʻia a i ʻole kolamu capillary ma muli o ka paila paila, polarity a me adsorption coefficient o kēlā me kēia ʻāpana i ka hāpana.ʻO ka ʻokoʻa, ua hoʻokaʻawale ʻia nā ʻāpana i ke kolamu, a laila ʻike ʻia nā mea ʻike e pili ana i ke kolamu e like me nā ʻano physicochemical o nā ʻāpana, a ma hope e hoʻouna ʻia nā hōʻailona uila i hoʻololi ʻia i ke kahua hana chromatographic.Hoʻopaʻa ʻia a hoʻopaʻa ʻia ka chromatogram kinoea o kēlā me kēia ʻāpana e kahi hale hana chromatographic e loaʻa ai nā hopena loiloi o kēlā me kēia ʻāpana.
●Hoʻohana ia i ka 10/100M adaptive Ethernet kamaʻilio kamaʻilio a kūkulu ʻia i loko o ka IP protocol stack, e hiki ai i ka mea kani ke hoʻomaopopo maʻalahi i ka lawe ʻana i ka ʻikepili mamao ma o ka LAN kūloko a me ka Pūnaewele.Hoʻomaʻamaʻa ia i ka hoʻonohonoho ʻana o ka hale hana a maʻalahi i ka hale hana.Hoʻomaʻamaʻa ka hoʻonohonoho ʻana i ka hoʻokele ʻana i ka ʻikepili analytical.
●Hoʻolālā o 3 kaʻina hoʻohui kūʻokoʻa.Hiki ke hoʻopili ʻia me ka hoʻoponopono kūloko (laboratory), ʻoi aku ka maʻalahi o nā luna e nānā i ke kūlana holo a nānā i nā hopena.
●Hiki i ke kahua hana NetChromTM ke kākoʻo i nā chromatographs he nui no ka hoʻoponopono ʻikepili a me ka hoʻomalu ʻana, e hoʻomaʻamaʻa i ka hoʻokele palapala a me ka hōʻemi ʻana i ka hoʻopukapuka lab mea hoʻohana a me nā lilo hana.
●Pūnaehana ma ka ʻōlelo Pelekania a me ka Pākē, hiki ke hoʻololi manuahi.
●Hiki ke kapa inoa ʻia e ka mea hoʻohana, kahi kūpono no nā mea hoʻohana.
●Ke hoʻohana nei i kahi ʻano hana like like multiprocessor, ʻoi aku ka paʻa a hilinaʻi.Hoʻopili ia i ka hoʻohālikelike laʻana paʻakikī.Hoʻolako ʻia me nā ʻano mea ʻike kiʻekiʻe e like me FID, TCD, ECD, a me FPD.Hiki ke hoʻokomo ʻia i ʻehā mau hoʻokolohua i ka manawa like.Hiki nō hoʻi ke hoʻohana i ke ʻano hoʻohui detector, kahi kūpono e kūʻai a hoʻokomo i nā mea ʻike ʻē aʻe ma hope o ke kūʻai ʻia ʻana o ka mea kani.
●Hoʻohana ka mea hana i ke ʻano modular, hoʻolālā ʻia a hoʻololi ʻia, maʻalahi e hoʻonui a pale i ka pono o ka hoʻopukapuka.
●ʻO ka ʻōnaehana hoʻomalu wela microcomputer hou he kiʻekiʻe ka hoʻomalu ʻana i ka wela a me ka hilinaʻi kiʻekiʻe a me ka hana anti-interference.Loaʻa iā ia nā ʻōnaehana hoʻomalu wela kūʻokoʻa kūʻokoʻa ʻeono, hiki ke hoʻomaopopo i ka papahana wela 16-step, kahi e hiki ai i nā mea hana ke hoʻohālikelike i ka laʻana.Loaʻa iā ia kahi ʻōnaehana wehe puka hope no ka umu, e hoʻomaikaʻi i ka pololei o ka mana haʻahaʻa haʻahaʻa a hoʻonui i ka wikiwiki o ka piʻi ʻana / hoʻohaʻahaʻa.
●Hiki ke hoʻolako ʻia ka mea kani me ka mea hoʻokele uila uila holomua (EFC) a me ka mea hoʻokele uila uila (EPC) e ʻike i ka mana kikohoʻe, hiki ke hoʻomaikaʻi nui i ka reproducibility o nā hopena qualitative a me quantitative.
●Hiki i ka polokalamu hoʻolālā hoʻolālā hoʻomaka ponoʻī ke hoʻopau maʻalahi i ka nānā ʻana ma ka pūnaewele o nā laʻana kinoea (koi i nā ʻāpana injection aunoa pūnaewele).
●ʻO ka ʻōnaehana hana o ka microcomputer control keyboard holoʻokoʻa he maʻalahi a maʻalahi hoʻi e hana;a ua hoʻolālā ʻia ka ʻenehana ʻike maka o ka mea ʻike.Me ka ʻike hewa a me ka pale ʻana i ka ʻikepili, hiki ke hoʻopaʻa ʻia nā ʻāpana.
●Hoʻokumu ka mea kani i ka leo haʻahaʻa, kiʻekiʻe kiʻekiʻe 24-bit AD kaapuni me ka waihona baseline a me ka unuhi kumu.
●E hōʻiliʻili i nā hōʻailona chromatographic a me ka hana ʻikepili, kūpono no WinXP, Win2000, Windows7 a me nā ʻōnaehana hana ʻē aʻe.Heluhelu ʻia ka ʻikepili i hoʻohālikelike ʻia mai nā faila CDF i kūpono i nā kūlana A/A (American Analytical Society), no laila hiki ke hoʻohana ʻia me Agilent, Waters a me nā wahi hana chromatography ʻē aʻe.
●ʻO ka ʻōnaehana Chromatographic me nā kuleana waiwai naʻauao kūʻokoʻa loa, me ka interface maʻamau o MODBUS/TCP, hiki ke maʻalahi me DCS.
●Hiki ke hoʻopili ʻia ka mea kani me ka sampler kaʻa i hana ʻia e nā mea hana hale a me nā ʻāina ʻē;e like me Shimadzu AOC-20i, HTA hui ka HT moʻo wai kiʻekiʻe-efficiency, kinoea mīkini sampler.
Nā kinoea ʻike | H2, CO, CO2, CH4, C2H4, C2H6, C2H2.(7 kinoea),O2(koe), N2 (koe) |
Dpāʻani | 192*64 kiko matrix LCD |
Wahi hoʻomalu mahana | 6 ala |
Palena hoʻomalu mahana | ka wela lumi +5℃~400℃, hoʻonui 1℃, pololei:±0.1℃ |
ʻO ke kauoha papahana | 16 kapuai |
Ka piʻi ʻana o ka papahana | 0.1~40℃/min |
Mana kinoea | ʻO ke ʻano hoʻomalu valve mechanical, ke ʻano hoʻomalu kahe uila uila |
Nā hanana kūwaho | 4 auwai |
ʻAno injector | kolamu i hoʻopaʻa ʻia, capillary, ʻeono ʻaoʻao valve kinoea injection, ʻakomi headspace injection, etc. |
Ka helu o nā mea ʻike | 4 pcs;FID, TCD, ECD, FPD (koho) |
E hoʻomaka i ka ʻinikini | manual, ʻakomi koho |
Pākuʻi kamaʻilio | Ethernet: IEEE802.3 |
Nā kikoʻī kikoʻī | |
ʻO ka mea ʻike ionization flame hydrogen (FID) | |
palena ʻike | ≤ 2 × 10-11g / s (n-hexadecane / isooctane); |
Ka leo kumu | ≤ 5 × 10-14A |
Ka neʻe ʻana o ka pae kumu | ≤ 1 × 10-13A / 30 min |
Laina laina | ≥106 |
Mea ʻimi hoʻoheheʻe wela (TCD) | |
ʻO ka ʻike | S≥2500mV•ml/mg (benzene/toluene) (hoʻonui 1, 2, 4, 8 manawa koho) |
Ka leo kumu | ≤20μV |
Ka neʻe ʻana o ka pae kumu | ≤30μV/30min |
Laina laina | ≥104 |
Mea ʻimi hopu uila (ECD) koho | |
palena ʻike | ≤ 1 × 10-13g / ml (propylene hexa-6 / isooctane) |
Ka leo kumu | ≤0.03mV |
Ka neʻe ʻana o ka pae kumu | ≤0.2mV/30min |
Laina laina | 103 |
Puna radioactive | 63Ni |
Mea ʻimi kiʻi kiʻi ahi (FPD) koho | |
palena ʻike | (S) ≤ 5×10-11g / s, (P) ≤ 1×10-12g/s;methyl parathion/ ʻāpala piha loa) |
Ka leo kumu | ≤ 3 × 10-13A |
Ka neʻe ʻana o ka pae kumu | ≤ 2 × 10-12A / 30 min |
Laina laina | Fa i ʻole sulfur ≥ 102, no ka phosphorus ≥ 103 |
Huina Kino Kino Kino | 1 hoʻonohonoho |
mīkini hana waihā | 1 hoʻonohonoho |
mīkini hana ea | 1 hoʻonohonoho |
ʻO Shaker nui-hana | 1 hoʻonohonoho |
ʻO ka pahu hoʻohaʻahaʻa kaomi maʻamau | 1 hoʻonohonoho |
Nitrogen Pressure Reducing Valve | 1 hoʻonohonoho |
Poʻo Nitrogen | 1 omole |
kinoea maʻamau | 1 omole |
Nati ala kinoea M8X1Φ3.2 | 10 pcs |
ʻO ka hui pahu hoʻohaʻahaʻa kaomi | 5 pcs |
Nut kolamu capillary Φ0.8 | 4 pcs |
Nut kolamu Φ4.2 a me Φ3.2 | ʻO kēlā me kēia 5 pcs |
ʻO ke kinoea kinoea kuhiliʻole i loko o ka paipu Φ2 | 10 pcs |
Papa hoʻopaʻa ala kinoea | 50 pcs |
ʻO ka pahu hoʻoheheʻe ʻana i ka capillary sampler evaporative pad | 50 pcs |
Kapili graphite pad Φ0.8 | 20 pcs |
Papa graphite Φ4.2 a me Φ3.2 | ʻO kēlā me kēia 10 pcs |
Fuse | 4 pcs |
Screw M3X5 | 6 pcs |
Paipu kinoea PE Φ3 | 20m |
1ml mea hoʻoheheʻe plastik | 5 pcs |
100ml injector | 5 pcs |
Needle | 10 pcs |
20ml aniani injector | 5 pcs |
5ml aniani injector | 5 pcs |
1ml aniani injector | 5 pcs |
Paʻi nui | 2 pcs |
Vaseline 30g | 2 pcs |
50mm 70mm | ʻO kēlā me kēia 1 pc |
50mm 70mm | ʻO kēlā me kēia 1 pc |
Uku 8-10 | 1 pc |
Uku 12-14 | 2 pc |
Paʻa ihu ʻoi | 1 pc |
uwea honua | 1 pc |
Uea hōʻailona | 1 pc |
Uila mana | 1 pc |
Kāpena lāʻau | 20 pc |
ʻeke ʻeke alumini | 2 pcs |
nila pālua | 25 pcs |
Disk | 1 pc |
Uea kamaʻilio | 1 pc |
Palaka kinoea | 1 pc |